New TopShim run:

version = topshim 2.0.0
date = 2023-11-21
time = 20:51:24.714
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.2

initial B0 stdDev = 3.34 Hz
sample size = 2.77 cm, position = 0.06 cm
final B0 stdDev = 0.41 Hz > improvement = 8.1
envelope width = 0.92 Hz

Performing tune after shimming
tune lock level increase = 1.1

shim changes: 
Z	+28
Z2	-47
Z3	+112
Z4	-310
Z5	+48
X	-51
Y	+137

duration = 3 min 39 sec
completed successfully
finished Tue Nov 21 20:55:02 2023


========================================================================